Remove Now
Cancle

Automatic Biochemistry Analyzer JH-9050

Multi-standard Analysis, Linearity & Nonlinearity Calibration Single & Dual reagents can be used

Description

System Function

  • Type: Full automatic , random access

  • Though put: 400tests/hour, 800test with ISE

  • Assay tapes: End point, Kinetics, fixed time, immunoturbidimetry, Monochromatic, Bi-chromatic & Multi-standard Analysis Linearity & Nonlinearity Clibra
    Single & Dual Reagents can be used

  • Test Range: Clinical chemistry , Specific problem, ISE, Drug abuse, Urine, Cerbrospinal fluid,  etc.

 

Sample System

  • Sample tray: 115 sample position

  • Sample Volume: 1.5 ~35μl (in steps of 0.1μl)

  • Sample probe: clot detection, liquid level detection

  • Dilution function

 

Reagent System

  • Reagent Tray: 90 reagent positions.

  • Reagent volume: 15μl~300μl

  • Barcode reader is optional

 

Reaction System

  • Reaction tray: 120 cuvette light pass length: 6mm

  • Reaction liquid volume: 150μl~300μl

  • Reaction temperature: 37℃±0.1℃

     

 

 

Optical System

  • Light source: Halogen lamp 6V/10W, light span≥3000 hours

  • Grating Measuring wavelength: 340 to 800nm, 12 wavelengths invariable

  • Detector: photodiode array

  • Rear spectrophotometry of holographic concave flat field grating

 

Calibration and Quality control

  • Assay tapes: End point, Kinetics , fixed time. immunoturbidimetry,  Monochromatic,Bi-chromatic &

    Multi-standard Analysis, Linearity & Nonlinearity Calibration Single & Dual reagents can be used

  • Calibration Mode: Linear (one-point, two point and mult-point), logit-log4p, logit-log5p nSpine, Exponential , Polynomial ,Parabola.

  • Calibration Period: setting as required.

  • OC Rule: Westgard Multi-rule, Cam μ Lative sum check and Twin plot

  • OC Mode: Intra day & Inter day

 

Working Condition:

  • System interface: Network port

  • Power Supply : 220V AC.50/60Hz

  • Power consumption: ≤1000VA

  • Environmental temperature: 15~30℃

  • Humidity: ≤85%

  • Dimension:1146X760X1095mm

  • Weight: 500kg





Automatic Biochemistry Analyzer JH-9050

Year Established:2013

Business Type:

Management Certifiation:

Operational Address: Facade Room, Linsheng Mansion, Gaogang Dist., Taizhou, Jiangsu, China (Mainland)