Multi-function Desktop Calibrator
High-precision, large range, bipolar, low drift desktop calibrator.
Wide dynamic range of source
Description
High-precision, large range, bipolar, low drift desktop calibrator.
Wide dynamic range of source
Voltage +/- 30.000V Current +-220.00mA High resolution: 1uV/100nA
Multi-type thermocouple support: T J E K S R B N
Multi-type communication support remote control: RS232, USB, WIFI(option)
Software support
Vacuum fluorescent display VFD
Scan mode/ Program mode
Memory and recall 10 group setting.
Floating, bipolar output system.
VFD Display:5digital
Power supply :110/220VAC 50/60Hz
.png)
Applications:
Calibration service
Companise service
Instrument and control workshops
Industry (laboratories,workshops and production)
Quality assurance
Specification:
Voltage +/- 30.000V/3.0000V/300.00mV/30.000mV,
Voltage source range
| Range |
Source range |
Setting resolution |
| 30mV |
0 to ±32.000mV |
1μV |
| 300mV |
0 to ±320.00mV |
10μV |
| 3V |
0 to ±3.2000V |
10μV |
| 30Vv |
0 to ±32.000V |
1mV |
Voltage source
| Range |
Overall accuracy
(1 year) |
Overall accuracy
(90 days) |
1-day stability
(23℃±1℃) |
1-day stability
(23℃±5℃) |
| ±(% of setting +V) |
| 30mV |
0.03+1μV |
0.025+4μV |
0.006+4μV |
0.01+4μV |
| 300mV |
0.03+25μV |
0.025+20μV |
0.006+5μV |
0.01+10μV |
| 3V |
0.025+200μV |
0.02+200μV |
0.005+20μV |
0.008+30μV |
| 30V |
0.025+2 mV |
0.02+2 mV |
0.005+100μV |
0.008+20μV |
Current +/- 220.00 mA /30.000 mA /3.0000 mA
Current source range
| Range |
Source range |
Setting resolution |
| 3mA |
0 to ±3.2000 mA |
100nA |
| 30mA |
0 to ±32.000 mA |
1μA |
| 200mA |
0 to ±220.00 mA |
10μA |
Current source
| Range |
Overall accuracy
(1 year) |
Overall accuracy
(90 days) |
1-day stability
(23℃±1℃) |
1-day stability
(23℃±5℃) |
| ±(% of setting +A) |
| 3mA |
0.03+250nA |
0.025+250nA |
0.008+20nA |
0.01+20nA |
| 30mA |
0.03+2.5μA |
0.025+2.5μA |
0.008+200nA |
0.01+200nA |
| 200mA |
0.035+25μA |
0.03+25μA |
0.008+2μA |
0.01+2μA |
Thermocouple temperature T J E K S R B N -200~1800 degC
Resolution 0.1
| Thermocouple |
Source range |
Setting resolution |
| T(CC) |
-200.0℃ to +400.0℃ |
0.1℃ |
| J(IC) |
-200.0℃ to +1200.0℃ |
0.1℃ |
| E(CRC) |
-200.0℃ to +1000.0℃ |
0.1℃ |
| K(CA) |
-200.0℃ to +1372.0℃ |
0.1℃ |
| S(PR10) |
-10.0℃ to +1768.0℃ |
0.1℃ |
| R(PR13) |
-10.0℃ to +1768.0℃ |
0.1℃ |
| B(PR30) |
+330.0℃ to +1820.0℃ |
0.1℃ |
| N |
-200.0℃ to +1300.0℃ |
0.1℃
|
Shanghai Yi Hua V&A Instrument Co., Ltd.